    Common faults and causes of zirconia oxygen analyzers

    發布時間: 2024-12-13  點擊次數: 903次

    1. System maintenance.

    If the oxygen indicator shows zero or no oxygen indication in the oxygen quantity converter, first check whether the oxygen indication circuit open. If the circuit is normal, check whether the zirconia probe is normal. At a temperature of 700℃, measure the DC millolt value of the oxygen cell. If the potential is greater than 1 millivolt and the resistance is greater than 100Ω, it indicates the probe is aged and needs to be replaced with a sensor or probe. If the potential is less than 1 millivolt and the resistance is less than100Ω, it indicates an open circuit in the wiring, which should be checked and handled one by one.

    2. constant temperature system failure.

    (1) If the heating temperature cannot be maintained, check whether there is a short circuit or open circuit in the resistance wire inside the heating furnace The normal resistance of the heating furnace wire is about 60-80Ω. If the resistance wire is open or the resistance measured by the multimeter less than 60Ω, the heating furnace wire should be replaced.

    (2) When the converter shows that the temperature is too high or out of control, it not only causes inaccurate oxygen display but also acceler the aging or damage of the zirconia head. At this point, you should immediately check if the thermocouple is open-circuited, or a multimeter to check the thermocouple. If the resistance is greater than 20Ω, it is judged that the thermocouple is open-cuited, and the thermocouple and zirconia head should be replaced; another reason for the temperature失控is: relay or temperature control circuit failure, relay or temperature control circuit failure will cause the oxygen converter to always be in one of the heating voltages of the heating furnace, thus leading to an increase in.

    3. oxygen cell fault.

    (1) After the oxygen cell has been used for a period of time, the contact points of its leads may become oxidized, resulting in contact or detachment. This can cause the Em value (in units of a multimeter) to fluctuate and become unstable. Observe the oxygen cell and take action.

    (2) When the oxygen signal deviation is too large or inaccurate, it may be due to a short circuit in the zirconia electrode, which case the wiring of the zirconia cell should be checked for short circuits. When the temperature does not rise, if the internal resistance is less than 0Ω, it indicates a short circuit in the electrode, and the zirconia head needs to be replaced.

    (3) During the operation of the oxygen measurement device, if the background potential exceeds 10mV or is less than -30mV, it must be recalibrated. If the background potential remains abnormal after calibration, it indicates that the probe has aged and can no longer be used.

    4. pneumatic circuit failure.

    (1) The zirconia head is clogged with ash, causing an obstruction in the reference gas circulation. This results in a decrease in probe output potential Em and an increase in the background potential Eo. In this case, blowing several times with a bulb syringe at the gas inlet of the will resolve the issue after the gas line is unclogged.

    (2) Oxygen cell ash blockage: The probe should be removed from the ash blockage and the filter should be reinstalled.

    (3) Probe leakage has a significant impact on the probe.

    5. The oxygen content on both sides of the boiler is too high.

    By adjusting the boiler combustion conditions, when starting or stopping the boiler fan or coal mill group, if the oxygen content deviation on both sides of the boiler is large, it is a normal phenomenon. Once the boiler combustion conditions stabilize, the oxygen content display will return to normal. If the unit adjustment is not in place and the deviation is large on both sides, it is necessary to check if the zirconia head is working properly by dismantling the transmitter for inspection. The device requires the operator to adjust the gas flow. The gaseous and reference state gases should continuously flow through the electrode without stopping, and there should be no dead ends. However, too large a flow will have a cooling effect, making the cell temperature uneven and causing errors. If the flow is too small, it is not only easy to form stagnant gas flow and dead zones but also may cause local changes in oxygen concentration due to the conduction of oxygen ions, thus affecting the accuracy of the measurement and causing deviations on both sides.

    6. On-site interference causes failures.

    For newly installed instruments, due to the strong electromagnetic field, when an unstable fault occurs, the first thing to consider is the interference issue.


在線客服 聯系方式 二維碼

服務熱線

0550-2409223

掃一掃,關注我們

皖公網安備 34118102000850號

亚洲AV永久无码精品一区二区国产 | 国产V片在线播放免费无码| 精品久久久久久久久中文字幕| 蜜桃成人无码区免费视频网站| 最近中文字幕2019高清免费| 天堂无码久久综合东京热| 亚洲精品无码久久千人斩| 大地资源中文第三页| 亚洲AV无码乱码在线观看牲色| 亚洲精品无码MV在线观看| 线中文在线资源 官网| 中文字幕乱妇无码AV在线| 岛国av无码免费无禁网| 久久精品aⅴ无码中文字字幕不卡 久久精品aⅴ无码中文字字幕重口 | 最近中文字幕mv免费高清在线| 无码任你躁久久久久久| 免费精品无码AV片在线观看| 国产成人无码AⅤ片在线观看| 色欲狠狠躁天天躁无码中文字幕| 国产精品99无码一区二区| 男人的天堂无码动漫AV| 无码人妻AV一二区二区三区| 中文字幕AV中文字无码亚| 无码人妻久久一区二区三区蜜桃| 中文字幕av一区| 中文字幕你懂的| 精品人妻V?出轨中文字幕| 中文字幕一区二区三区在线观看| 亚洲精品中文字幕乱码三区| 超清中文乱码字幕在线观看| 高清无码中文字幕在线观看视频| 亚洲一区二区三区无码影院| 乱人伦中文无码视频在线观看| 中文无码久久精品| 今天免费中文字幕视频| 久久久久综合中文字幕| 蜜桃无码一区二区三区| 亚洲精品高清无码视频 | 日韩人妻精品无码一区二区三区| 亚洲AV综合色区无码一区爱AV | 人妻少妇乱子伦无码视频专区|